Nanmac LogoNanmac Logo

ISO 9001-2015

ISO 17025:2017 (Calibration)

AS 9100D by Q-4 2020

Service Disabled Veteran Owned Small Business

MADE IN USA

Contact Us

About UsProductsCalibrationUltra Fast (Eroding) ThermocouplesCertificationsWhitepapersNews
Certifications / ISO/IEC 17025:2017 Accreditation
logo

PERRY JOHNSON LABORATORY ACCREDITATION, INC.

Certificate of Accreditation

Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:

Nanmac Corp.

1657 Washington Street, Holliston, MA 01746

(Hereinafter called the Organization) and hereby declares that Organization is accredited in accordance with the recognized International Standard:

ISO/IEC 17025:2017

This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (as outlined by the joint ISO-ILAC-IAF Communiqué dated April 2017):

Calibration of Thermocouples and RTD sensors

(As detailed in the supplement)

Accreditation claims for such testing and/or calibration services shall only be made from addresses referenced within this certificate. This Accreditation is granted subject to the system rules governing the Accreditation referred to above, and the Organization hereby covenants with the Accreditation body’s duty to observe and comply with the said rules.

For PJLA:

logo

Tracy Szerszen

President/Operations Manager

Perry Johnson Laboratory Accreditation, Inc. (PJLA) 755 W. Big Beaver, Suite 1325 Troy, Michigan 48084

Initial_Accreditation_Date

November 15, 2016

Issue Date:

November 18, 2018

Expiration Date:

February 28, 2021

Accreditation No.:

89989

Certificate No.:

L18-541

The validity of this certificate is maintained through ongoing assessments based on a continuous accreditation cycle. The validity of this certificate should be confirmed through the PJLA website: www.pjlabs.com

logo

Certificate of Accreditation: Supplement

Nanmac Corp.

1657 Washington Street, Holliston, MA 01746 Contact Name: Michael Pike Phone: 508-872-4811

Accreditation is granted to the facility to perform the following calibrations:

Thermodynamic

MEASURED INSTRUMENT, QUANTITY OR GAUGERANGE OR NOMINAL DEVICE SIZE AS APPROPRIATECALIBRATION AND MEASUREMENT CAPABILITY EXPRESSED AS AN UNCERTAINTY (±)CALIBRATION EQUIPMENT AND REFERENCE STANDARDS USED
Temperature Measurement Thermocouple Type B F800 °C to 1 700 °C 1.5 °C Hart 9009
Pico TC-08
Pico PT104
Temperature Measurement Thermocouple Type C F0 °C to 1 150 °C
0.7 °C
1 150 °C to 1 700 °C
1.5 °C
Fluke 9150
DAQ USB-TEMP
Fluke 1586A
Temperature Measurement Thermocouple Type D F0 °C to 1 150 °C
0.7 °C
1 150 °C to 1 700 °C
1.5 °C
Type S reference
Type_B_reference
RTD_Reference
Temperature Measurement Thermocouple Type E F0 °C to 500 °C
500 °C to 900 °C
0.4 °C
0.6 °C
Camco Furnace
Saturn Furnace
Harper Furnace
Temperature Measurement Thermocouple Type J F0 °C to 500 °C
0.4 °C
500 °C to 750 °C
0.6 °C
Temperature Measurement Thermocouple Type K F0 °C to 500 °C
0.4 °C
500 °C to 800 °C
0.6 °C
800 °C to 1 372 °C
1.4 °C
Temperature Measurement Thermocouple Type N F0 °C to 500 °C
0.4 °C
500 °C to 1 150 °C
0.6 °C
1 150 °C to 1 372 °C
1.4 °C
Temperature Measurement Thermocouple Type R F0 °C to 500 °C
0.6 °C
500 °C to 800 °C
0.8 °C
800 °C to 1 450 °C
1.5 °C
Temperature Measurement Thermocouple Type S F0 °C to 500 °C
0.7 °C
500 °C to 800 °C
0.9 °C
800 °C to 1 450 °C
1.5 °C
Temperature Measurement Thermocouple Type T F0 °C to 400 °C0.4 °C
RTD Sensors F0 °C to 300 °C
0.37 °C
300 °C to 600 °C
0.94 °C
Hart 9009
Pico PT-104
Fluke 9150
RTD_Reference
Fluke 1586A
Saturn Furnace
  • 1. The CMC (Calibration and Measurement Capability) stated for calibrations included on this scope of accreditation represents the smallest measurement uncertainty attainable by the laboratory when performing a more or less routine calibration of a nearly ideal device under nearly ideal conditions. It is typically expressed at a confidence level of 95 % using a coverage factor k (usually equal to 2). The actual measurement uncertainty associated with a specific calibration performed by the laboratory will typically be larger than the CMC for the same calibration since capability and performance of the device being calibrated and the conditions related to the calibration may reasonably be expected to deviate from ideal to some degree.
  • 2. The laboratories range of calibration capability for all disciplines for which they are accredited is the interval from the smallest calibrated standard to the largest calibrated standard used in performing the calibration. The low end of this range must be an attainable value for which the laboratory has or has access to the standard referenced. Verification of an indicated value of zero in the absence of a standard is common practice in the procedure for many calibrations but by its definition it does not constitute calibration of zero capacity.
  • 3. The presence of a superscript F means that the laboratory performs calibration of the indicated parameter at its fixed location. Example: Outside Micrometer F would mean that the laboratory performs this calibration at its fixed location.

Issue 11/2018

This supplement is in conjunction with certificate #L18-541

ContactIndustries ServedTechnical LiteratureCapabilitiesTerms & Conditions

© 2020 Nanmac Corporation

Follow us onNanmac LinkedIn Page Link